14C Contamination Testing in Natural Abundance Laboratories: A New Preparation Method Using Wet Chemical Oxidation and Some Experiences
Abstract
Substances enriched with radiocarbon can easily contaminate samples and laboratories used for natural abundance measurements. We have developed a new method using wet chemical oxidation for swabbing laboratories and equipment to test for 14C contamination. Here, we report the findings of 18 months’ work and more than 800 tests covering studies at multiple locations. Evidence of past and current use of enriched 14C was found at all but one location and a program of testing and communication was used to mitigate its effects. Remediation was attempted with mixed success and depended on the complexity and level of the contamination. We describe four cases from different situations. Show more
Permanent link
https://doi.org/10.3929/ethz-b-000123364Publication status
publishedExternal links
Journal / series
RadiocarbonVolume
Pages / Article No.
Publisher
Cambridge University PressSubject
Contamination; Wet chemical oxidiation; Preparation; 14C; RecoveryOrganisational unit
08619 - Labor für Ionenstrahlphysik (LIP) / Laboratory of Ion Beam Physics (LIP)
Notes
It was possible to publish this article open access thanks to a Swiss National Licence with the publisher.More
Show all metadata
ETH Bibliography
yes
Altmetrics