An 11k-electrode 126-channel high-density microelectrode array to interact with electrogenic cells
Abstract
A microelectrode array allows an arbitrary group of 126 electrodes to be selected from a total of 11,016 in order to do cell or neural recordings from areas of interest with 18 mum spatial resolution and 2.4 μv input-referred noise. Signals are amplified by 0 to 80dB, bandpass filtered (0.3 to 4kHz), and finally digitized (20kS/s, 8b). Example recordings from acute brain slices are shown. Show more
Publication status
publishedExternal links
Book title
2007 IEEE International Solid-State Circuits ConferenceJournal / series
Digest of Technical Papers / IEEE International Solid State Circuits ConferenceVolume
Pages / Article No.
Publisher
IEEEEvent
Organisational unit
03684 - Hierlemann, Andreas / Hierlemann, Andreas
More
Show all metadata
ETH Bibliography
yes
Altmetrics