Al2O3:C optically stimulated luminescence dosimeters (OSLDs) for ultra-high dose rate proton dosimetry

Open access
Author
Show all
Date
2021-04-21Type
- Journal Article
Citations
Cited 13 times in
Web of Science
Cited 16 times in
Scopus
ETH Bibliography
yes
Altmetrics
Abstract
The response of Al2O3:C optically stimulated luminescence detectors (OSLDs) was investigated in a 250 MeV pencil proton beam. The OSLD response was mapped for a wide range of average dose rates up to 9000 Gy s−1, corresponding to a ~150 kGy s−1 instantaneous dose rate in each pulse. Two setups for ultra-high dose rate (FLASH) experiments are presented, which enable OSLDs or biological samples to be irradiated in either water-filled vials or cylinders. The OSLDs were found to be dose rate independent for all dose rates, with an average deviation <1% relative to the nominal dose for average dose rates of (1–1000) Gy s−1 when irradiated in the two setups. A third setup for irradiations in a 9000 Gy s−1 pencil beam is presented, where OSLDs are distributed in a 3 × 4 grid. Calculations of the signal averaging of the beam over the OSLDs were in agreement with the measured response at 9000 Gy s−1. Furthermore, a new method was presented to extract the beam spot size of narrow pencil beams, which is in agreement within a standard deviation with results derived from radiochromic films. The Al2O3:C OSLDs were found applicable to support radiobiological experiments in proton beams at ultra-high dose rates. Show more
Permanent link
https://doi.org/10.3929/ethz-b-000483996Publication status
publishedExternal links
Journal / series
Physics in Medicine and BiologyVolume
Pages / Article No.
Publisher
Institute of PhysicsSubject
FLASH; proton dosimetry; optically stimulated luminescence; Al2O3:C; ultra-high dose rateMore
Show all metadata
Citations
Cited 13 times in
Web of Science
Cited 16 times in
Scopus
ETH Bibliography
yes
Altmetrics