Intracorneal Ring Segment Implantation Results in Corneal Mechanical Strengthening Visualized With Optical Coherence Elastography
OPEN ACCESS
Loading...
Author / Producer
Date
2022-07
Publication Type
Journal Article
ETH Bibliography
yes
Citations
Altmetric
OPEN ACCESS
Data
Rights / License
Abstract
PURPOSE:
To quantify the mechanical impact of intracorneal ring segment (ICRS) implantation of different dimensions in an ex vivo eye model.
METHODS:
A total of 30 enucleated porcine eyes were assigned to ICRS implantation (thickness: 300 µm, angle: 120°, 210°, or 325°), tunnel creation only, or virgin control groups. For mechanical evaluation, each globe was mounted on a customized holder and intraocular pressure (IOP) was increased in steps of 0.5 mm Hg from 15 to 17 mm Hg, simulating physiologic diurnal IOP fluctuations. At each step, an optical coherence tomography volume scan was recorded. Deformations between subsequent scans and the locally induced axial strains were analyzed using a vector-based phase difference method. The effective E-modulus was derived from the overall induced strain as a measure of global mechanical impact.
RESULTS:
ICRS implantation increased the effective E-modulus from 146 and 163 kPa in virgin and tunnel-only eyes to 149, 192, and 330 kPa in eyes that received a 5-mm optical zone ICRS with 120°, 210°, and 325° arc length, respectively; and to 209 kPa in a 6-mm optical zone ICRS with 325° arc length. The most consistent effect was a shift toward positive strains in the posterior stroma by 0.1% to 0.46% (factor 1.15 to 2.15) after ICRS surgery.
CONCLUSIONS:
ICRS implantation reduces the overall tissue strain under the load of the IOP and provokes posterior tissue relaxation. This effect is more prominent the longer the arc length and the smaller the optical zone of the ICRS is. ICRS have not only a geometrical, but also a mechanical impact on corneal tissue. This behavior might have clinical implications when ICRS implantation is performed in biomechanically weakened keratoconic corneas.
Permanent link
Publication status
published
External links
Editor
Book title
Journal / series
Volume
38 (7)
Pages / Article No.
459 - 464
Publisher
SLACK
Event
Edition / version
Methods
Software
Geographic location
Date collected
Date created
Subject
Organisational unit
Notes
Funding
174113 - Measuring local corneal biomechanical properties by multi-frequency vibrography: Moving towards an earlier diagnosis of pathologies and personalized computer simulations (SNF)