Commissioning and quality assurance of a novel solution for respiratory-gated PBS proton therapy based on optical tracking of surface markers
OPEN ACCESS
Loading...
Author / Producer
Date
2022-02
Publication Type
Journal Article
ETH Bibliography
yes
Citations
Altmetric
OPEN ACCESS
Data
Abstract
We present the commissioning and quality assurance of our clinical protocol for respiratory gating in pencil beam scanning proton therapy for cancer patients with moving targets. In a novel approach, optical tracking has been integrated in the therapy workflow and used to monitor respiratory motion from multiple surrogates, applied on the patients’ chest. The gating system was tested under a variety of experimental conditions, specific to proton therapy, to evaluate reaction time and reproducibility of dose delivery control. The system proved to be precise in the application of beam gating and allowed the mitigation of dose distortions even for large (1.4 cm) motion amplitudes, provided that adequate treatment windows were selected. The total delivered dose was not affected by the use of gating, with measured integral error within 0.15 cGy. Analysing high-resolution images of proton transmission, we observed negligible discrepancies in the geometric location of the dose as a function of the treatment window, with gamma pass rate greater than 95% (2%/2 mm) compared to stationary conditions. Similarly, pass rate for the latter metric at the 3%/3 mm level was observed above 97% for clinical treatment fields, limiting residual movement to 3 mm at end-exhale. These results were confirmed in realistic clinical conditions using an anthropomorphic breathing phantom, reporting a similarly high 3%/3 mm pass rate, above 98% and 94%, for regular and irregular breathing, respectively. Finally, early results from periodic QA tests of the optical tracker have shown a reliable system, with small variance observed in static and dynamic measurements.
Permanent link
Publication status
published
External links
Editor
Book title
Journal / series
Volume
32 (1)
Pages / Article No.
52 - 62
Publisher
Elsevier
Event
Edition / version
Methods
Software
Geographic location
Date collected
Date created
Subject
Proton therapy; Moving tumours; Respiratory gating; Optical tracking; Surface markers; Commissioning; Quality assurance